Chapter 6 _______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 183
Calibration procedure finished.
Calibration: FAILED.
Zero test: ok (0.0000, limit:
absolute +/-0.0001)
Signal test: fail (52.4083, limits:
recalibration +/-0.0300 rejection +/-
0.2000)
Remove calibrator assembly and type
YES: YES
1>
Guided Calibration
Procedure failed due to
Signal Value Test failure.
Signal Value deviation is
measured to 52.4%. Limit for
rejection is ±0.2 (±20%).
Table 34 below shows an example of a Guided Calibration Procedure
where the Zero Signal Test (and therefore the Calibration Procedure)
fails due to a too high receiver signal with blocked transmitter and
receiver.
Table 34 Guided Calibration Procedure, Zero Test Failed
0> LEVEL 1
Operating level set to: 1
1> CALIBRATE FS11
Step 1/3
Clean Windows and type YES: YES
Step 2/3
Install Zero Plugs and type YES:
YES
1st run
..........||||||||
Zero signal test FAILED. (3.7833)
Verify that zero plugs are in
place.
Repeat the test? Type YES: YES
2nd run
.........|||||||||
Zero signal test FAILED. (3.7829)
Verify that zero plugs are in
place.
Repeat the test? Type YES: YES
3rd run
.................|
Zero Signal Test is failing
three times in a row. The
most common handling
failure (no Zero Plugs
installed) is mentioned as
hint to succeed the test.
The user has to confirm
the restart of the test.
Calibration procedure finished.
Calibration: FAILED.
Zero test: fail (3.7829, limit:
absolute +/-0.0001)
Guided Calibration
Procedure failed due to Zero
Signal Test failure. Zero
Value is measured to 3.78.