The DRY OFFSET value must be between 700 and 900, when the
RAINCAP
®
hardware operates normally. The DRY command without a
parameter just shows this parameter.
WET
The WET command without a parameter shows the scaling factor that
normalizes the RAINCAP
®
signal change from the dry state to the wet
state to be 1.00. A typical value is close to 0.0015.
DRD WET SCALE 0.00169
The WET ON command is used to set the parameter. The RAINCAP
®
measuring surfaces must be coated with a wet cloth or immersed in water
for at least 30 seconds, when the WET ON command is given. Make
sure the wet cloth touches only the RAINCAP
®
sensor.
The wet scale normalizes the measured signal range (DRY - WET) to
1.00.
HEAT
This test command can be used to test the heater control relays and the
heating elements and power.
The following is an example of the command:
HEAT ON
When you want to stop the heater test and quit the test mode, press ESC.
NOTE
Do not leave the heater testing on in warm conditions for long periods of
time.
Other Commands
TIME
The TIME command is used for maintenance purposes. To display the
current system time, type the following:
TIME
The following is an example of the system output:
10:11:12