Chapter 6 _______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 185
9. The signal value must be close to the one printed on the plates. If
the difference is less than 5%, calibration is correct. If not, continue
with the calibration procedure.
10. Terminate the CHEC command by pressing the ESC key.
0308-006
Figure 50 Assembling Opaque Glass Plates