Chapter 6 _______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 179
12. Confirm the entered signal value by typing YES and Enter or re-
enter the signal value after typing NO and Enter.
- The Scatter Signal Test starts and the progress is indicated on
the screen.
- The test step can take up to two minutes.
- When the Scatter Signal Test is finalized a test result summary
and a passed/ fail indication for the Scatter Signal Test is given.
Step 3/3
Remove zero plugs, install glass plates and enter
signal value (between 0 and 10): 1.3497
Is signal value 1.349700 correct?
Type YES or NO? YES
1st run
..............||||
Signal test PASSED. (1.3514)
Calibration procedure finished.
Calibration: PASSED.
Zero test: ok (0.0000, limit: absolute +/-0.0001)
Signal test: ok (0.1259, limits: recalibration +/-
0.0300 rejection +/-0.2000)
Remove calibrator assembly and type YES: YES
1>
13. Remove the calibrator assembly and the calibrator arm and store
them in the case of the Calibrator Kit.
14. Type YES and press enter.