EasyManua.ls Logo

Vaisala HMT370EX - Page 78

Vaisala HMT370EX
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1. Attach the welding joint to the process pipe or chamber.
2. Apply a sealant (MEGA-PIPE EXTRA No. 7188 or LOCTITEâ No. 542 with activator No.
7649) on the threads of the welding joint and screw the bottom of the ball valve onto the
welding joint.
3. Tighten the ball valve assembly by turning from the ball valve body.
Tightening the ball valve kit by turning the extension nipple
can break the sealing. Tighten the ball valve assembly only from the ball
valve body.
CAUTION!
4. If you need to cap the ball valve assembly before installing or after removing the probe,
attach a blanking nut to close the top of the valve.
HMT370EX User Guide M212305EN-B
76

Table of Contents

Related product manuals