User's Guide ______________________________________________________________________  
152 _________________________________________________________________ M210667EN-C 
Command Description 
 SET INTERNAL_FILTER number  Sets filter value of the internal filter. When an internal 
filter is applied (for example for installations with baseline 
length of less than 30 m), it is necessary to set the 
internal filter value accordingly. Otherwise the displayed 
SIGNAL values in Alignment Menu and Calibration Menu 
will be incorrect. 
Example: Configure an internal filter with 56 % 
transmission respectively an attenuation factor of 0.56 
with the command SET INTERNAL_FILTER 0.56 
  SET MESSAGE SECOND 
 TYPEmessage_number 
Selects the message type to be transmitted as the 
Second Message. 
  SET MESSAGE SECOND TYPE MITRAS  Selects the MITRAS Single Base emulation mode for the 
Second Message. The data message and message 
frames are the same as they are in Vaisala MITRAS 
Transmissometer. 
  SET MESSAGE SECOND TYPE 
 FD12MSG2 
Selects the FD12 message 2 emulation mode for the 
Second Message. The data message and message 
frames are the same as they are in Vaisala FD12 
Visibility Meter. 
  SET MESSAGE SECOND TYPE 
 FD12MSG7 
Selects the FD12 message 7 emulation mode for the 
Second Message. The data message and message 
frames are the same as they are in Vaisala FD12P 
Weather Sensor. 
  SET MESSAGE SECOND TYPE SKOPO2  Selects the SKOPOGRAPH II Flamingo emulation mode 
for the Second Message. The data message and 
message frames are the same as they are in 
Impulsphysik SKOPOGRAPH II Flamingo. 
  SET MESSAGE SECOND PORT DATA   The Second Message is directed to the Data port.  
  SET MESSAGE SECOND PORT 
 MAINTENANCE 
The Second Message is directed to the Maintenance 
port.  
  SET MESSAGE SECOND PORT MODULE   The Second Message is directed to the optional 
communication module, for example, modem module.  
  SET MESSAGE SECOND PORT NONE   No Second Message is given to any port (default). 
  SET METAR OFF  Disables METAR messaging (default). 
  SET METAR ON  Enables METAR messaging. without framing 
  SET METAR 8  Enables METAR messaging. with framing 
  SET PWD OFF  Disables PWD data polling in the LT31 system. 
  SET PWD ON  Enables PWD data polling in the LT31 system. 
  SET PWD_MOR_ONERROR [ON|OFF]  The current status is reported by set pwd_mor_on_error. 
The response is PWD MOR is (NOT) used on error. 
To disable the use of PWD MOR in error situations use 
the command set pwd_mor_on_error off. The response 
would be PWD MOR is NOT used on error. 
To enable the use of PWD MOR in error situations 
(default) use the command set pwd_mor_on_error on. 
The response is PWD MOR is used on error. 
 SET PWD_PITCH number  During auto-calibration, derived PWD MOR data will be 
corrected by a factor with the value of number. To take 
PWD data as it is, number has to be set to 1.000 
(default). 
  SET MESSAGE FRAME OFF  Message frames not included in data messages. 
  SET MESSAGE FRAME ON  Message frames included in data message transmission 
(default). 
 SET MOR_RANGE minimum_MOR 
 maximum_MOR 
Under and overrange values for MOR readings are set. 
Values are to be given in meters. The absolute minimum 
value is 10. The absolute maximum value is 10 000. The 
minimum value is set according to the baselength. A 
baselength up to 35 m will allow underrange values 
down to 1/3 of the baselength. A baselength above 35 m 
will allow underrange values down to 1/2 of the 
baselength.