User's Guide ______________________________________________________________________  
184 _________________________________________________________________ M210667EN-C 
Original LT31 messages from 1 to 9 are unchanged. The visibility data of 
PWD22 message 7 (one minute and ten minutes averages) is invalidated 
by slashes. 
The unit identifier given in the PWD22 message 7 is the same as set for 
LT31 unit identifier. When the unit identifier for LT31 is removed, the 
unit identifier given in the PWD22 message 7 is 0 (zero, ASCII-0x30). 
There is no crosscheck for whether the message interval and port settings 
are able to offer a bandwidth large enough to transmit all the data (for 
example, 300 bps, 15 s, LT31 message 9, METAR-Messaging enabled 
does not work). 
 
System Configuration 
The SET command is used to set or update system communication and 
user interface related parameters. 
The CALIBRATE command is used to set or update visibility, window 
contamination and temperature measurement calibrations. This command 
is available only at the advanced level. For calibration instructions, see 
section Calibration on page 232. 
The current system parameters can be displayed by using the 
PARAMETERS command. Type the following: 
 0> PARAMETERS