Chapter 6 ______________________________________________________________ Maintenance 
VAISALA ______________________________________________________________________ 249 
Calculated values and deviation are updated continuously and the 
deviation can be read (see Figure 144 below). 
 
 
  0401-125 
Figure 144  Linearity Test Menu, Filter Applied 
The applied filter value is given. The calculated values are displayed as 
269.8 m for MOR and 71.6732 % for transmittance. The deviation of the 
current measurement compared to the calculation can be read by the 
means of the values for Deviation (abs.) and (rel.) for MOR and 
transmittance. 
NOTE 
Take into account that all measured values are 1-minute-averages. After 
applying an optical filter, the system needs its response time to handle 
the new situation. 
 
NOTE 
The 2-minutes-Variation-versus-current-Measurement is given only 
when the latest action (entering the menu page, applying the filter, 
removing the filter) occurred longer than two minutes ago. Otherwise it 
is replaced by dashes ('---------'). 
The Reference-Quality-Indication is given only when, by procedure, no 
filter is applied and the latest action (entering the menu page, removing 
the filter) occurred longer than two minutes ago. Otherwise it is replaced 
by dashes ('--------'). See Figure 145 on page 250. 
 
If any other optical filter will be applied or the linearity test is finished, 
the user has to remove the currently applied filter first. 
Remove Filter After Measurement! Done? (yes/no): 
To confirm, type yes<CR>.  
To abort the action, type 
no<CR> .