Chapter 6 ______________________________________________________________ Maintenance 
VAISALA ______________________________________________________________________ 215 
CAUTION 
Handle the module carefully. Do not touch the components on the board 
to avoid ESD damage to them. Ground yourself before attempting to 
touch the board. 
 
CAUTION 
Do not let the module get wet. 
 
5.  Replace the LTL212 and tighten the fastening screws carefully. 
Secure the cables with a cable tie. 
6.  The spare LTL212 module is fully adjusted and tested, no 
additional adjustments are necessary. 
7.  After replacing the LTL212 module, pay attention to the correct 
position of the optics unit inside the measurement unit enclosure. It 
should be mounted in parallel to the enclosure walls. 
8.  Reconnect the three plugs at the rear end of the LTC112 board. 
9.  Tighten the two fastening screws and the bolt carefully. 
10.  Close the measurement unit carefully. Pay attention to the position 
of the cover seal. 
11.  The repaired measurement unit must be mounted in the same way 
as described in sections Mounting Receiver Measurement Unit on 
page 124 and Mounting Transmitter Measurement Unit on page 
114. 
12.  Make sure that the enclosure is parallel to the support unit.  
13.  Secure the fastening screws of the measurement unit carefully. 
14.  Turn on the AC power switch (and also the battery backup switch, 
if installed). 
CAUTION 
After repairing the measurement unit, fine alignment and calibration 
must be carried out to establish normal operational conditions. For 
details, see sections Fine Alignment on page 231 and Calibration on 
page 232. 
 
15.  Re-install the measurement head cover and lock it carefully. 
16.  Close the interface unit and mount the upper interface unit cover 
after fine alignment and calibration is finalized.