User's Guide ______________________________________________________________________  
264 _________________________________________________________________ M210667EN-C 
 
C2:Transmitter 
Measurement CPU 
LTC112; window 
receiver AC 
overrange  
Window receiver signal is greater 
than overrange threshold while 
window transmitter LED driver is 
detected as faulty. 
Check derating and intensity value for 
window transmitter module (wTdera, wTint) 
(see section Service on page 250). 
Repl
ace window transmitter module (see 
section Replacing Window Transmitter 
Module  on p
age 213). 
C3:Transmitter 
Measurement CPU 
LTC112; window 
receiver AC 
overrange  
Window receiver signal is greater 
than overrange threshold while 
window transmitter LED driver is 
detected as OK. 
Replace transmitter measurement CPU 
(see section Replacing Measurement 
CPU/
Window Receiver  on page 211). 
C5:Receiver 
Measurement CPU 
LTC112; window 
receiver AC 
overrange  
Window receiver signal is greater 
than overrange threshold while 
window transmitter LED driver is 
detected as faulty. 
Check derating and intensity value for 
window transmitter module (wTdera, wTint) 
(see section Service on page 250) 
Repl
ace window transmitter module (see 
section Replacing Window Transmitter 
Module  on p
age 213).  
C6:Receiver 
Measurement CPU 
LTC112; window 
receiver AC 
overrange  
Window receiver signal is greater 
than overrange threshold while 
window transmitter LED driver is 
detected as OK. 
Replace receiver measurement CPU (see 
section Replacing Measurement 
CPU/
Window Receiver  on page 211). 
L1:LM21; indication 
of abnormal situation  
LM Sensor has detected a hood 
or dew heater problem. 
Refer to LM21 User's Guide. 
L2:LM21; alarm -> 
BGL not valid  
LM Sensor has detected at least 
one of the following situations: 
- the sensor does not react to 
  changes in luminance 
- field calibrator has been applied
- backscatter signal has increased
Refer to LM21 User's Guide. 
L3:LM21; error -> 
BGL not valid  
LM21 Sensor has detected a 
memory error 
Refer to LM21 User's Guide. 
L4:LM21; not 
responding  
LM21 sensor does not respond to 
LT31 system polling request 
Check whether the LM Sensor lines (power, 
RS–485) are connected properly (see 
section Mounting Optional LM21 on page 
98). 
LM sen
sor is not foreseen at all, but LM 
Sensor is configured to be available (refer 
to Table 11 on page 150). 
L5:LM21; not 
installed  
LM Sensor is configured to be not 
available (refer to Table 11 on 
page 150). 
–