Chapter 7 ___________________________________________________________ Troubleshooting 
VAISALA ______________________________________________________________________ 269 
 
R34:Receiver Unit 
LTR111; not 
responding 
Receiver Unit does not respond 
to LT31 system polling request 
Check whether all the Receiver Unit lines 
are connected properly 
Check whether all connections concerning 
the RS-485 Module Bus are connected 
properly. 
R35:Receiver Unit 
LTR111; unknown  
Any Receiver Unit Module has 
been changed since last Offset 
and Visibility Calibration. 
Perform an Offset and Visibility Calibration 
(see section Offset and Visibility Calibration 
on pag
e 234). 
T1:Transmitter Main 
Transmitter LTL112; 
LED driver error  
T2:Transmitter Main 
Transmitter LTL112; 
LED driver error  
The intensity value of the Main 
Transmitter LED is out of limits for 
normal operation or has been 
evaluated as to be off. 
T3:Transmitter Main 
Transmitter LTL112; 
LED driver error  
 
T4:Transmitter Main 
Transmitter LTL112; 
LED driver error  
The derating value of Main 
Transmitter LED has been 
evaluated as to be off. 
Replacement of the Main Transmitter 
Module is necessary (see section 
Replacing Main Transmitter Module  on 
page 216). 
T5:Transmitter Main 
Transmitter LTL112; 
LED aged  
The derating value of the Main 
Transmitter LED has been 
evaluated as to be high. 
Replacement of the Main Transmitter 
Module in mid–term range (see section 
Replacing Main Transmitter Module  on 
page 216). 
T6:Transmitter Main 
Transmitter LTL112; 
LED error  
The derating value of the Main 
Transmitter LED has been 
evaluated as to be too high. 
Replacement of the Main Transmitter 
Module (see section Replacing Main 
Tran
smitter Module  on page 216). 
T7:Transmitter 
Window Transmitter 
LTL212; LED driver 
error  
T8:Transmitter 
Window Transmitter 
LTL212; LED driver 
error  
The intensity value of the Window 
Transmitter LED is out of limits for 
normal operation or has been 
evaluated as to be off. 
T9:Transmitter 
Window Transmitter 
LTL212; LED driver 
error  
 
T10:Transmitter 
Window Transmitter 
LTL212; LED driver 
error  
The derating value of Window 
Transmitter LED has been 
evaluated as to be off. 
Replacement of the Window Transmitter 
Module is necessary (see section 
Replacing Window Transmitter Module  on 
page 213).