Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
109 / 188
COPYRIGHT© 2018 VATECH.
ALL RIGHTS RESERVED
7) Set the parameters as shown below and click CONFIRM.
8) Click READY when enabled and acquire the image according to the Console Software
guidance and save the image in a 2D Viewer (EzDent-I or EasyDent 4).
9) Open the saved Phantom image in the 2D Viewer to run the 3D Viewer (Ez3D-i or Ez3D plus).
10) Go to the Axial view in full-screen mode by clicking (for Ez3D plus) or by double-clicking
(for Ez3D-i).
11) Click > (for Ez3D plus)
12) Click Measurement > ROI (for Ez3D-i)