EasyManua.ls Logo

Waters 600e - Page 42

Default Icon
258 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14 Basic Operation
Powering Up a 600E System When Used with a Data System
This section contains the powerup sequence to follow when using the 600E system with a
data system (Empower or Millennium
®
Chromatography Manager).
Note: The 600 Controller must be in the Gradient Controller configuration to be controlled
from an IEEE-488 based data system. See Section 2.4.2, Gradient Controller
Configuration, for details.
Note: Power up all devices connected to your data system before you power up the
computer. As you power up each component, wait a brief period to allow its internal
diagnostic tests to run. These tests ensure that each module is functional and serve to
quickly isolate a failure.
To power up the 600E system when used with a data system:
1. Power up all equipment not controlled by the data system (for example, a printer).
2. Power up all equipment controlled by the data system but not under direct IEEE-488
control (for example, a non-IEEE-488 detector):
3. Power up all equipment controlled through the IEEE-488 bus.
To power up the 600E system, press the On/Off switch on the front panel of the 600
controller to the 1 (On) position. The Powerup screen appears (Figure 2-1).
Figure 2-1 Gradient Controller Powerup Screen
The Powerup screen lists the diagnostic status, controller configuration, and software
version number.

Table of Contents

Other manuals for Waters 600e

Related product manuals