Setting Up the 600E Pump 69
To determine the flow factor for your application:
1. Set the flow rate at 1.0 mL/min.
2. Measure the actual flow rate.
3. If the actual flow rate is less than or greater than 1.0 mL/min, type a flow factor
value to compensate for the difference. The valid range is 0.950 to 1.050. For
example, if your actual flow rate is 0.98 mL/min, use a flow factor of 1.02.
4. Press Enter.
Autostart
Specifies a time delay before a run starts and tells the system where to begin execution in
a Method Table (step number and vial number). You can use Autostart at any time.
While the autostart countdown is in progress, you can continue to run a method started
before you initiated Autostart. When the Autostart countdown completes, the system
immediately aborts any run in progress, and the Autostart parameters take precedence.
To Autostart a method:
1. Press the Set Up function key to display the Pump Setup screen.
2. Move the cursor to the Autostart field.
3. Type the total delay time in the Hrs and Min fields and press Enter.
Enter a value between 0 and 99 for Hrs (hours).
Enter a value between 0 and 59 for Min (minutes).
4. At the Step field, type the step number of the Method Table where program
execution will start. Press Enter.
5. At the Vial field, type the vial number in the specified step number where sample
analysis will start. Press Enter.