Settings inoLab
®
Multi 9620 IDS
106 ba77173e04 12/2017
System / General / Temperature
unit
°C
°F
Temperature unit,
degrees Celsius or degrees Fahren-
heit.
All temperature values are displayed
with the selected unit.
System / General / Barometric
pressure unit
mbar
mmHg
inHg
Air pressure unit
System / General / Stability control on
off
Switches on or off the automatic stabil-
ity control during measurement (see
section 11.7.3 A
UTOMATIC STABILITY
CONTROL, page 107 )
System / Interface / Baud rate 1200, 2400, 4800,
9600, 19200
Baud rate of the USB-B interface (USB
Device)
System / Interface / Output format ASCII
CSV
Output format for data transmission
For details, see section 13 T
RANSMIT-
TING DATA, page 118
System / Interface / Decimal sepa-
rator
Dot (xx.x)
Comma (xx,x)
Decimal separator
System / Interface / Output header Output of a header for Output format:
CSV
System / Interface / Extended Oxi
output
The measured values for the parame-
ters concentration (mg/l ) and satura-
tion (%) are output together.
The function is active if the following
requirements are met:
a D.O. sensor is connected
The D.O. sensor displays the
parameter concentration (mg/l) or
saturation (%)
the Output format CSV is set
System / Clock function Date format
Datum
Time
Settings of time and date.
For details, see section 4.5.5 E
XAMPLE
2
ON NAVIGATION: SETTING THE DATE
AND TIME, page 25
System / Service information Hardware version and software ver-
sion of the meter are displayed.
System / Reset - Resets the system settings to the
default values.
For details, see section 11.8.2 R
ESET-
TING THE SYSTEM SETTINGS, page 111
Menu item Setting Explanation