EasyManua.ls Logo

wtw inoLab Multi 9620 IDS - Page 51

wtw inoLab Multi 9620 IDS
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
inoLab
®
Multi 9620 IDS Ion concentration
ba77173e04 12/2017 51
Continuing with two-
point calibration
8. Thoroughly rinse the electrode with distilled water.
9. Immerse the electrode in standard solution 1.
10. When calibrating without temperature sensor:
Measure the temperature of the standard solution using a thermom-
eter.
Use <F4>/[ ] to select the setting of the temperature.
Use <><> to set the temperature.
Use <F4>/[ ] to select the setting of the concentration.
11. Set the concentration of the standard solution with <><> and press
<ENTER>.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
12. Wait for the end of the AutoRead measurement or accept the calibration
value with <ENTER>.
The calibration display for the next standard solution appears.
01.09.2017 08:00
01.09.2017 08:00
Standard
13. Thoroughly rinse the electrode with distilled water.
14. Immerse the electrode in standard solution 2.

Table of Contents

Related product manuals