Multi 9620 IDS Ion concentration
ba77173e04 12/2017 65
Measurement
7. Start the measurement with <F1> Continue
An entry window appears.
8. Prepare the standard solution.
9. If necessary, add the ISA/TISAB solution to the standard solution.
10. Thoroughly rinse the electrode with deionized water.
11. Immerse the electrode in the standard solution (complexing agent or
precipitating agent).
12. Using <><> and <ENTER>, enter the values for the volume of the
standard solution (Std. volume), the concentration of the standard solu-
tion (Std. conc.) and the volume of the ISA/TISAB solution (ISA/TISAB
vol.).
13. Select Continue with <><> and start the measurement with
<ENTER>.
When the measurement is finished an entry window appears.
14. Add the sample to the standard (complexing agent or precipitating
agent).
15. Using <><> and <ENTER>, enter the value for the volume of the
sample (Sample volume).
16. Select Continue with <><> and start the measurement with
<ENTER>.
The measurement result is displayed when the measurement is com-
pleted.
i
Immerse sensor in stan-
dard
Std. volume 100.0 ml
Std. conc. 1.0 mg/l
ISA/TISAB vol. 1.0 ml
Continue
Sample subtraction
01.09.2017 08:00
i
Add sample!
Sample volume 1.0 ml
Continue
Sample subtraction
01.09.2017 08:00

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw inoLab Multi 9620 IDS and is the answer not in the manual?

wtw inoLab Multi 9620 IDS Specifications

General IconGeneral
Brandwtw
ModelinoLab Multi 9620 IDS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals