inoLab
®
Multi 9620 IDS Conductivity
ba77173e04 12/2017 85
Calibration evalua-
tion
After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Calibration record
(example)
Display Calibration record Cell constant [cm
-1
]
+++ Within the range
0.450 ... 0.500 cm
-1
Error Error Outside the range
0.450 ... 0.500 cm
-1
Error elimination (see section 15
W
HAT TO DO IF..., page 125)
CALIBRATION Cond
01.09.2017 07:43:33
TetraCon 925
Ser. no. 09250033
Cell constant 0.476 1/cm
25.0 °C
Sensor +++