5.11 Averaging Times Setting
Procedure
1.
Press SETUP. The soft key menu for sweep condition settings appears.
2.
Press the AVERAGE TIMES soft key. The averaging times setting screen is
displayed.
3.
Enter a number of averaging times using the rotary knob, arrow keys, or numeric
key pad.
4.
Press ENTER. The value specified by the AVERAGE TIMES soft key is
displayed.
Note
Increasing the averaging times reduces the sweeping speed but the S/N improves.
Explanation
AVERAGE TIMES
This key is used to set the average times for each point.
The setting range is 1 to 999.
If you press the COARSE key you can use the rotary knob or arrow keys to change
the numerical value in 1-2-5 steps. The value changes in 1 step if you do not press the
COARSE key.
Note
When the sweep speed is set to 2x, times are not averaged. The number of averaging times
can be set when the sweep speed is set to 1x.