Analysis in the Zoom Area
Power Measurement in the Zoom Area
Optical power can be measured by integrating the measured waveform level 
measurements in the zoom area. 
1.
  Zoom in on the measured waveform. Set the range you want to measure to the 
display scale. For the procedure on zooming in on the waveform, see section 6.1, 
“Zooming In/Out on a Waveform.”
2.
  Press the MARKER switch. The soft key menu regarding the markers appears. 
3.
  Press the MORE soft key repeatedly until the MORE 3/3 menu is displayed.
4.
  Press the SEARCH/ANA ZOOM AREA soft key, and select ON. 
     
5.
 Press ANALYSIS. The soft key menu for analyzing measured waveforms 
appears. 
6.
  Press the ANALYSIS 1 soft key. The analysis function selection menu is 
displayed. 
7.
  Press the POWER soft key. Analysis is performed between display scales, and 
the results are displayed in the data area.
   
Note
•  When the zoom area search function is enabled,   is displayed in inverse video.
•  If the SEARCH/ANA ZOOM AREA key is OFF, analysis is performed over the entire range of 
the measurement scale.
•  For details on the POWER analysis function, see appendix 3, “Details of Analysis 
Functions.” 