9-24
IM AQ6370D-01EN
Turning the Display OFF
This function can be used to temporarily turn the display OFF. Use this function when 
you are using the instrument in a darkroom or similar environment in which the light from 
the screen has an effect on the work.
DISPLAY OFF
Press this soft key to turn OFF the backlight, which will also turn the display OFF.
When Remote Commands Have Been Used to Turn the Display OFF
If remote commands have been used to turn the display OFF, even if you perform panel 
key or mouse operations to turn the display ON, the following message will be displayed 
and the display will be turned OFF after approximately 5 seconds.
   
To keep the display turned ON, use a remote command, or press LOCAL to switch the 
instrument from remote to local mode.
Clearing Measured Data, Analysis Conditions, Parameters, Etc.
This function clears all measurement parameters, display parameters, analysis 
parameters, and the waveform display. It is the same operation as when the AQ6370D 
receives an *RST remote command.
The following data is not cleared.
•  Remote interface (ETHERNET, GP-IB, and RS232C) settings
•  Wavelength calibration data and alignment adjustment data
•  Various type of data saved to internal memory
•  Registered programs created using the program function
9.6  Other Settings