EasyManuals Logo

YOKOGAWA AQ6370D User Manual

YOKOGAWA AQ6370D
445 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #418 background imageLoading...
Page #418 background image
App-47
IM AQ6370D-01EN
Appendix
1
2
3
4
5
6
7
8
9
10
11
App
Index
XdB_STOP-BAND
λ
i
Wavelength
Level [dB]
nominal wavelength
x
dB
xdB stop-band
Port
A
Port B
Parameter
THRESH
Procedure
Obtain the width (xdB_stop-Band) that centers on the nominal wavelength λi of each
channel and that is located below by the parameter THRESH_LEVEL both on the left
and on the right.
XdB_ELIMINATION BAND
λ
i
Wavelength
nominal wavelength
Port
A
Port B
Level [dB]
xdB Elimnation
band
xdB
Test Band
Maximum Level
in Test Band
Parameter
THRESH
TEST BAND
Procedure
• Perform the peak search within the range of the parameter Test_Band/2 by centering
on the nominal wavelength λi of a channel and obtain the peak level (LPi).
• Obtain the width (xdB_Elimination_Wavelength) that is above the peak level LPi
obtained in the above procedure by the parameter THRESH_LEVEL both on the left
and on the right.
Appendix 6 Details of Optical Filter Analysis Function

Table of Contents

Other manuals for YOKOGAWA AQ6370D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA AQ6370D and is the answer not in the manual?

YOKOGAWA AQ6370D Specifications

General IconGeneral
BrandYOKOGAWA
ModelAQ6370D
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals