EasyManua.ls Logo

Zeiss Axio Examiner - Page 59

Zeiss Axio Examiner
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPERATION
Axio Examiner Illumination and contrast procedures Carl Zeiss
M60-2-0003 e 05/2012 59
(3) Setting of a transmitted-light bright field
according to KÖHLER
The Axio Examiner has been put into
operation properly (see chapter 3).
The transmitted-light illumination is switched
on at the external power supply unit
(Figure 4-1/1).
Set image brightness using the light intensity
controller (Figure 4-1/2) at the external power
supply unit.
Place a high-contrast specimen in the universal
mounting frame on the mechanical stage.
Swing the front lens at the condenser in
position (for objectives 10x), and set the
condenser to the upper stop using the pinion
knob for height adjustment (Figure 4-2/2). The
stop must be set to ensure that the specimen
will not be lifted out of position by the
condenser.
Switch the revolving disk of the condenser
(where existing) using the knurled ring
(Figure 4-2/4) to the H position for the bright
field.
Swing the 10x objective in position at the
objective mount or objective nosepiece
(Figure 4-2/7), and focus on the specimen using
the pinion knob (Figure 4-2/3).
Close the luminous-field diaphragm
(Figure 4-2/5) until the luminous-field
diaphragm also becomes visible in the visual
field (even if out of focus) (Figure 4-2/A).
Lower the condenser using the pinion knob for
height adjustment until the bright field stop rim
appears sufficiently in focus (Figure 4-2/B).
Center the luminous-field diaphragm using
both centering screws (Figure 4-2/1) at the
condenser carrier (Figure 4-2/C), and then open
it until the stop rim just disappears from the
visual field (Figure 4-2/D).
Figure 4-1 HAL 100 external power supply unit
Figure 4-2 Microscope settings on the Axio
Examiner in the transmitted-light
bright field

Table of Contents

Related product manuals