EasyManua.ls Logo

Abbott CELL-DYN 3200 - Page 321

Abbott CELL-DYN 3200
684 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CELL-DYN
®
3200 System Operator’s Manual
6-23
9140181H—October 2001
Section 6 Calibration Procedures
Figure 6.8 Results Display Screen
As the first sample is processed, the [START AUTO-CAL] key disappears and the
[DELETE A RUN] key appears. The [ACCEPT] key appears after the results of the fifth
run are displayed.
The following information is displayed in the lower portion of the screen:
Current Factor
Mean Factor (new Calibration Factor)
Factor % Diff
•CV %
The Current Factor is calculated for each run. The Mean Factor is an average based
on all runs. The Factor % Diff shows the difference between the Mean Factor and
the existing Calibration Factor for that parameter. The CV (Coefficient of Variation
expressed as a percent) measures the degree of variation between runs.
NOTE: The Mean Factor is the new Calibration Factor.
If a result is highlighted in Green on the Run line, it indicates that parameter has
failed a reference or precision check. That result will not be included in the Mean
Factor and Factor % Diff calculations for that parameter.
Consequently, the Mean Factor for some parameters may be based on fewer runs
than other parameters and may even be based on fewer than 5 runs.

Table of Contents

Related product manuals