CELL-DYN
®
3200 System Operator’s Manual
6-39
9140181H—October 2001
Section 6 Calibration Procedures
9. To determine the New Calibration Factor for each parameter:
a. For Calibrator, use the values on the assay sheet and the CELL-DYN
mean values determined in steps 2 through 8 above. Enter this
information in columns 1 and 2 respectively of Worksheet 1, Open Mode
Calibration — New Factors, provided in Manual Calibration
Worksheets at the end of this section.
b. For Fresh Whole Blood, use the Reference Mean Values determined in
Determining Reference Values Fresh Whole Blood described earlier
(refer to the Whole Blood Calibration Reference Values Worksheet) and
the CELL-DYN Mean values determined in steps 2 through 8 above.
Enter this information in columns 1 and 2 respectively of Worksheet 1,
Open Mode Calibration — New Factors, provided in Manual
Calibration Worksheets at the end of this section.
c. Enter the current Open Sampler Calibration Factors, obtained in step 3
above, in column 3 of Worksheet 1.
NOTE: Be sure to make copies of this worksheet as needed.
10. Follow the instructions on Worksheet 1 to calculate the new Open Mode
Calibration Factor for each parameter and enter this information in column 4
of the worksheet. The method for determining the new factors is:
a. Calibrator Calibration:
b. Whole Blood Calibration:
For example, if the Reference Mean Value for WOC is 6.6, the CELL-DYN
Mean for WOC is 7.1, and the current Open Mode Calibration Factor for
WOC is 0.98, then:
(6.6 / 7.1) × 0.98 = 0.91
and 0.91 is your New Open Mode Calibration Factor for WOC.
Determining Which Parameters Need Calibration
To determine which parameters require calibration in the Open Mode, follow the
procedure below:
1a. Transfer the values from column 4 of Worksheet 1 to the new Open Mode
Factor columns 1 and 3 of Worksheet 2, Open Mode Factor % Difference,
provided in Manual Calibration Worksheets at the end of this section.
NOTE: Make copies of this worksheet as needed.
1b. Transfer the values for the Current Open Mode Cal factor from column 3 of
Worksheet 1 and enter these values in column 2 of Worksheet 2.
2. Follow the instructions on this worksheet to determine the Factor % Diff for
each parameter.
$VVD\9DOXH
&(//'<10HDQ
---------------------------------------------
&XUUHQW2SHQ0RGH
&DOLEUDWLRQ)DFWRU
×
1HZ2SHQ0RGH
&DOLEUDWLRQ)DFWRU
5HIHUHQFH0HDQ
&(//'<10HDQ
---------------------------------------------
&XUUHQW2SHQ0RGH
&DOLEUDWLRQ)DFWRU
×
1HZ2SHQ0RGH
&DOLEUDWLRQ)DFWRU