10-46
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
Flow sequence time out <x,x>/BL
NOTE: Characters in the brackets identify the problem flow sequence number
and name.
––– -Probable Cause–––– ––– -Corrective Action–––
1. An internal Analyzer flow sequence
was not completed in the allotted
time.
1. Record the characters in the
brackets.
2. Initialize the Analyzer by pressing
the [INITIALIZATION] key on the
second DIAGNOSTICS MENU
screen. Continue processing
samples.
3. If the problem recurs, contact your
Abbott Hematology Customer
Support Center.
Command sent at incorrect time <x>/BL
––– -Probable Cause–––– ––– -Corrective Action–––
1. The Analyzer received a command
from the System Software at the
incorrect time and could not process
it.
1. Record the characters in the
brackets.
2. Initialize the Analyzer by pressing
the [INITIALIZATION] key on the
second DIAGNOSTICS MENU
screen. Continue processing
samples.
3. If the problem recurs, contact your
Abbott Hematology Customer
Support Center.