CELL-DYN
®
3200 System Operator’s Manual
10-51
9140181H—October 2001
Section 10 Troubleshooting and Diagnostics
HSSL failure – bad outgoing commands/BL
The Analyzer received unrecognizable messages from the Data Module.
– - Probable Cause – - Corrective Action
1. The communication line has become
noisy, possibly due to a partial break
in the line.
1. Initialize the Analyzer by pressing
the [INITIALIZATION] key on the
second DIAGNOSTICS MENU
screen. If the message appears
again, power OFF the Analyzer.
Then reseat both ends of HSSL
cable and make sure the
connections are secure. Power
Analyzer ON.
NOTE: For the location of the
HSSL connectors on the
rear panel of the Data
Module, refer to Section
1: Use or Function;
Subsection: Data Module
Components.
NOTE: If you wish to use the
Sample Loader, reset the
racks before initializing or
rebooting the System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)