EasyManua.ls Logo

Abbott CELL-DYN 3200 - Page 486

Abbott CELL-DYN 3200
684 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10-52
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
HSSL failure – receiver hardware error/BL
The HSSL receiver on the Data Module malfunctioned.
– - Probable Cause – - Corrective Action
1. A hardware failure on the HSSL
card has occurred.
1. Initialize the Analyzer by pressing
the [INITIALIZATION] key on the
second DIAGNOSTICS MENU
screen. If the message appears
again, power OFF the Analyzer.
Then reseat both ends of HSSL
cable and make sure the
connections are secure. Power
Analyzer ON.
NOTE: For the location of the
HSSL connectors on the
rear panel of the Data
Module, refer to Section
1: Use or Function;
Subsection: Data Module
Components.
NOTE: If you wish to use the
Sample Loader, reset the
racks before initializing or
rebooting the System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Monitor power-on self-test error/BL
NOTE: This message pertains to a software execution monitor on the Analyzer,
not the video display monitor.
A failure occurred during self-test.
(e.g., test pattern not successfully read from memory)
– - Probable Cause – - Corrective Action
1. A failure in the Analyzer hardware
has occurred.
1. Reboot the System.
NOTE: If you wish to use the
Sample Loader, reset the
racks before rebooting the
System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)

Table of Contents

Related product manuals