Section 10 Troubleshooting and Diagnostics
Missing download program/BL
Download program not present on Data Module.
– - Probable Cause – - Corrective Action
1. Installation error, damaged hard
disk, or erased hard disk.
1. Reboot the System.
NOTE: If you wish to use the
Sample Loader, reset the
racks before rebooting the
System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Missing flow sequence/BL
Flow sequence not present on Data Module.
– - Probable Cause – - Corrective Action
1. Installation error, damaged hard
disk, or erased hard disk.
1. Reboot the System.
NOTE: If you wish to use the
Sample Loader, reset the
racks before rebooting the
System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Invalid substitutable parameter/BL
Size or structure of parameter (e.g., gain) is invalid.
– - Probable Cause – - Corrective Action
1. A software error has occurred. 1. Reboot the System.
NOTE: If you wish to use the
Sample Loader, reset the
racks before rebooting the
System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)