Section 14 Reticulocyte Package
Fault Report
When the [FAULTS REPORT] key is pressed, information regarding any pending
fault is displayed on the screen. The screen displays the words OPERATOR
CORRECTABLE FAULT REPORT: or FATAL FAULT REPORT and any
additional information available. If there is no fault, the screen displays the words
NO FAULT PENDING. Operator-correctable faults (for example, Waste Full,
Diluent Empty) can be cleared after the corrective action has been taken by
pressing the [CLEAR FAULT] key.
Execution Times
Display the time in tics used for various cycles during routine function of the
system.
NOTE: A tic is an integer that equals 0.055 seconds.
Retic Count Rate
Figure 14.23 Reticulocyte Count Rate Screen (Tabular Format)
When the [RETIC CNT RATE] key is pressed followed by [RETIC CNT DATA], the
following soft key labels are displayed:
RETIC CNT GRAPH
PRINT
RETURN
RETIC
CNT GRAPH
DIAGNOSTICS MENU
Ready
Aug 02 2001
Operator ID
Retic Seq#
14:32
R0482
TS
RETURNPRINT
RETIC: TOTAL COUNT: 31693
TIME:0.501.001.502.002.503.003.504.01
COUNT: 2181 4204 6348 8458 10517 12705 14832 16936
RATE: 4362.00 4046.00 4288.00 4220.00 4118.00 4376.00 4254.00 4166.34
TIME:4.515.015.516.026.527.027.50
COUNT: 19161 21247 23395 25507 27596 29692 31693
RATE: 4405.94 4172.00 4296.00 4182.18 4178.00 4192.00 4125.77