CELL-DYN
®
3200 System Operator’s Manual
1-17
9140181H—October 2001
Section 1 Use or Function
Waste Sensor Connector
The waste-full sensor plug connects to the Waste Sensor Connector port. The
ground shield on the cable should be attached to the Ground Connector on the
rear panel. When the electrical sensor is tripped, the EXTERNAL WASTE FULL
message is generated and the READY status is inhibited until the situation is
corrected. The Analyzer interprets a disconnected plug the same way as a full waste
container. Therefore, if the waste is routed to a drain, the dummy plug (supplied in
the accessory kit) must be inserted in the connector.
Waste Outlet Tube Connector
This port is used to connect the waste outlet tube.
Reagent Inlet Panel
Diluent/Sheath Inlet Tube Connector
This color-coded (red) port is used to connect the diluent inlet tube with its
associated cap, sinker and label.
WBC Lyse Inlet Tube Connector
This color-coded (purple) port is used to connect the WBC lyse inlet tube with its
associated cap, sinker and label.
HGB Lyse Inlet Tube Connector
This color-coded (blue) port is used to connect the HGB lyse inlet tube with its
associated cap, sinker and label.
Power Supply Module
The Power Supply Module supplies the power to operate the Analyzer and Data
Module.