7352 Series Digital Multimeter Operation Manual
7.1.4 DC Current Measurement (DCI)
7-6
7.1.4 DC Current Measurement (DCI)
7.1.4.1 Ach DC Current Measurement (DCI-Ach)
Input
terminal
Range
Maximum display Resolution
Resistance
between
terminals
(A fuse
resistance is
included.)
Measurement accuracy *1
(% of reading + digits)
Temperature
coefficient
(ppm of reading
+ digits)/C
FAST
MED/
SLOW1, 2
FAST
MED/
SLOW1, 2
FAST MED SLOW1, 2
Auto-
zero
ON
Auto-
zero
OFF
mA-
COM A
2000 nA 1999.9 1999.99 100 pA 10 pA 11.5 k or
less
0.15+35 0.15+40 0.15+40 150+4 150+4
20 A 19.999 19.9999 1 nA 100 pA 0.1+7 0.1+7 0.1+5 100+1 100+1.6
200 A 199.99 199.999 10 nA 1 nA 102 or less 0.04+20 0.04+40 0.04+40 40+4 40+5
2000 A 1999.9 1999.99 100 nA 10 nA 0.04+2 0.04+7 0.04+5 40+1 40+1.6
20 mA 19.999 19.9999 1 A 100 nA 2 or less 0.04+20 0.04+40 0.04+40 40+4 40+5
200 mA 199.99 199.999 10 A1 A 0.05+2 0.05+7 0.05+6 50+0.6 50+3
2000 mA 1999.9 1999.99 100 A10 A0.6 or less 0.15+2 0.15+7 0.15+6 150+0.6 150+3
*1: For Auto-zero ON For Auto-zero OFF, 2 digits are added to the digit term.
Maximum allowable
applied current
Input protection Fuse replacement
Between mA and
COM A terminals
2000 mA 2 A/250 V fast-blow fuse which is com-
pliant with IEC60127 sheet1
On the rear panel