7352 Series Digital Multimeter Operation Manual
7.1.7 Temperature Measurement (TEMP)
7-13
7.1.7 Temperature Measurement (TEMP)
Maximum allowable applied voltage
Compliant thermocouple standard: JIS C1602
Cold junction compensation: Internal
7.1.8 Ach Diode Measurement-Ach
Other specifications are the same as the resistance measurement function.
Thermocouple
K(CA) T(CC)
Input terminal Range -50C to 1370C-50C to 400C
TEMP HI - LO Maximum display 13700 4000
Resolution 0.1C0.1C
Measurement accuracy 0.15%3C 0.15%3C
Between TEMP (HI/LO) and COM A terminals 200 Vpeak
Between TEMP (HI/LO) and COM B terminals 200 Vpeak
Between TEMP (HI) and TEMP (LO) terminals 36 Vpeak
Between TEMP (HI/LO) terminal and the chassis 500 V
Input
terminal
Range
Maximum display Resolution
Measurement
current
Measurement accuracy *1
(% of reading + digits)
Temperature
coefficient
(ppm of reading
+ digits)/C
FAST
MED/
SLOW1, 2
FAST
MED/
SLOW1, 2
FAST MED SLOW1, 2
Auto-
zero
ON
Auto-
zero
OFF
VHz-
COM A
2000 mV 1999.9 1999.99 100 V10 V 1 mA 0.014+2 0.014+5 0.014+3 15+0.25 15+1.5
*1: For Auto-zero ON For Auto-zero OFF, 2 digits are added to the digit term. In addition, an offset error, which is obtained by multiply-
ing (the resistance of the measurement cable + 0.3 ) by 1 mA, is added.