7352 Series Digital Multimeter Operation Manual
5.3.2 Auto-range
5-10
5.3.2 Auto-range
When Auto-range is set to ON, the measurement range of this instrument is automatically set to the opti-
mum range depending on the input voltage.
The judgment value used when the range is changed
Range DOWN When the displayed value is less than 10% of the displayed range (less than 200 V if
the DCV 1000 V range and 700 V range are both selected)
Range UP When the displayed value exceeds the maximum display
Auto-range is set to OFF under the following conditions.
• Frequency measurement
• Diode measurement
• Continuity measurement
• The 10 A range for the DC and AC current measurements
• Temperature measurement