EasyManuals Logo

ADCMT 7352A User Manual

ADCMT 7352A
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
7352 Series Digital Multimeter Operation Manual
5.9.2 Smoothing Calculation
5-26
4. Canceling the NULL calculation
The NULL calculation is canceled when any of the following conditions occur.
The key is pressed while the NULL calculation is performed.
The *RST command is executed.
5.9.2 Smoothing Calculation
1. Function
The smoothing function is used when noise is superimposed on the measurement signal. Because
measurement values of the specified number of times (smoothing count) are calculated by using the
moving averaging method, the dispersion of those measurement values becomes small.
The smoothing calculation equation is shown below:
Pressing performs the calculation and the SM indicator in the display section turns on con-
tinuously or blinks.
If is pressed again while the smoothing calculation is performed, the calculation is canceled
and the SM indicator turns off.
The SM indicator blinks from the smoothing start to the (N - 1)th measurement and turns on after the
Nth measurement.
The (N + 1)th data is displayed as below:
The average of the measured values, which have been acquired, is displayed from the smoothing start
to the (N - 1)th measurement.
NULL
Display = (Measured value 1 + , , , , , + Measured value N) / N
SM
SM
Display = (Measured value 2 + , , , , , + Measured value (N + 1)) / N

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADCMT 7352A and is the answer not in the manual?

ADCMT 7352A Specifications

General IconGeneral
BrandADCMT
Model7352A
CategoryMultimeter
LanguageEnglish

Related product manuals