EasyManuals Logo

Advanced Instruments OsmoTECH PRO Service Guide

Advanced Instruments OsmoTECH PRO
75 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
Chapter 4: Diagnostics
31
OsmoTECH PRO Multi-Sample Micro-Osmometer Service Guide
4.11 Logging instrument maintenance and
repair
All instrument users can log routine instrument cleaning
from this screen.
Administrators and Service users can also record when
preventative maintenance and instrument repairs are
performed.
All instrument maintenance information is saved to the
events database.
NOTE: When needed, Administrators can also set the
installation date INSTALLATION tab. Refer to
Setting the installation date for instructions.
Logging routine instrument cleaning
1. Press MAINTENANCE and log in.
2. Press the CLEANING tab.
3. Select Cleaned Sample Probe and/or Cleaned
Cooling Well if those tasks were completed
4. Press Save.
5. Press OK on the Cleaning completed dialog box.
6. Press Exit to return to the Diagnostics menu.
Logging preventative maintenance (Administrators
only)
Administrator-level users and Service account users can
record when any preventative maintenance is performed
and the service engineer performing the maintenance.
The name of the service engineer is required for all
preventive maintenance events.
NOTE: Refer to Maintenance and Service for instrument
maintenance recommendations and instructions.
To record a maintenance event, complete the following
steps.
1. Press MAINTENANCE and log in.
2. Press the PREVENTATIVE MAINTENANCE tab.
3. Select the Advanced Instruments Preventative
Maintenance checkbox to record the maintenance
event.
4. Press to enter the name of the Advanced
Instruments trained service engineer (required)
who performed the maintenance.
5. Press EDIT to type a Comment (required - up to 1024
characters) to enter a
full explanation of the
maintenance performed
.
6. Press Save.
The instrument saves the preventative maintenance
information to the events database and updates the
preventative maintenance date. The instrument will
display prompts starting 11months after this
preventative maintenance date as a reminder that
the next recommended maintenance is due.
NOTE: To set the installation date (and initial
preventative maintenance date) for a new
instrument, refer to Setting the installation
date.
7. Press Exit to return to the Diagnostics menu.

Table of Contents

Other manuals for Advanced Instruments OsmoTECH PRO

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments OsmoTECH PRO and is the answer not in the manual?

Advanced Instruments OsmoTECH PRO Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmoTECH PRO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals