Chapter 7: Troubleshooting
53
OsmoTECH PRO Multi-Sample Micro-Osmometer Service Guide
Error Possible cause/Suggested action
Out of range Low
Out of range High
Cause: The sample might be outside
the range of the instrument:
Out of range Low indicates the
measured reading is below
-10mOsm.
Out of range High indicates the
measured reading is above
2200mOsm.
Action:
Confirm good pipette technique
and pipette condition
Test a control such as Clinitrol.
Check for particulate matter in
the sample.
Test timeout
Cause: The instrument was unable to
complete the test within the allotted
time.
Action:
Check for particulate matter in
the sample.
Test a control such as Clinitrol to
see if the issue persists.
If the issue persists, the sample
might be outside the range of the
instrument.
Confirm sample block and cooling
block are engaging (see Testing
the sample motor)
Sample tube not
detected
Cause: The sample tube was not
detected in the turntable.
Action:
Place a sample tube in the
indicated position.
If the sample tube is present but
not detected, try another load
position.
7.2 Errors during calibration
Error Possible cause/Suggested action
Tubes missing Action: Verify that all samples
required for calibration are loaded on
the turntable.
Wrong calibrator:
0 mOsm/kg, 50
mOsm/kg, 850
mOsm/kg, or 2000
mOsm/kg
Cause:
Calibrator’s result is below/above
the expected range for the
calibrator group twice
The calibration set up incorrectly
Wrong calibration standard in the
wrong position
Action:
Follow the onscreen instructions to
set up calibration and retry
calibration (see Calibrating the
instrument).
Use the correct calibrator.
Positions 1-3 should consist of
850mOsm
Positions 4-6 should consist of
50mOsm
Positions 7-9 should consist of
either 0mOsm or 2000mOsm
Positions 10-12 should consist of
2000mOsm (if successful
0mOsm calibration completed)
Action: Recalibrate the instrument
(see Calibrating the instrument)
Calibrator out of
range:
0mOsm/kg,
50mOsm/kg,
850mOsm/kg, or
2000mOsm/kg
Cause: Too much variation in the
replicates for the calibrator group.
Action:
Confirm good technique and
pipette position.
Check for particulates in sample.
Recalibrate the instrument (see
Calibrating the instrument)
Run Data isn’t saved
in Results table
1. Cycle power on the instrument.
2. Select > Diagnostics > Events.
If the software and firmware versions
at power up are not recorded, then
the SD Card has become Read-Only.
Either the SD card or the complete
LCD Display assembly (p/n 222022R)
will have to be replaced. The SD card
is not easy to access, therefore only
Advanced Instruments-trained service
engineers should attempt to change
it.

Table of Contents

Other manuals for Advanced Instruments OsmoTECH PRO

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments OsmoTECH PRO and is the answer not in the manual?

Advanced Instruments OsmoTECH PRO Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmoTECH PRO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals