Appendix C: OPC-UA Features and Configuration
60
OsmoTECH PRO Multi-Sample Micro-Osmometer Service Guide
8. Select the Remote results method node
9. Enter a date/time in ISO 8061 format (yyyy-MM-
ddThh:mm:ss) value in the Results requested
argument for this method, and submit the method
call.
The method returns output arguments in a JSON list
of database result records since the date/time
entered.
The OPC user can check the Tubes Status at any
time to show a status of T (tested) A (available), E
(error), B (being tested), P (present ready for test),
or U (unknown).
The OPC user can send a Remote Events request at
any time to examine recent events— settings
changes, errors, instrument power up, maintenance,
or cleaning—saved in the database.
JSON output:
The output of the Results and Events requests are
strings in JSON format. Each record is in one json object
where the field names are abbreviated such that:
DS = Description
RS = Result
SI = Sample ID
TN = Tube Number (position)
TS = Timestamp
TY = Type of event
UI = User Id

Table of Contents

Other manuals for Advanced Instruments OsmoTECH PRO

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments OsmoTECH PRO and is the answer not in the manual?

Advanced Instruments OsmoTECH PRO Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmoTECH PRO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals