Chapter 5: Configuring settings
45
OsmoTECH PRO Multi-Sample Micro-Osmometer Service Guide
Administrators can allow or restrict Supervisors from
performing all of the operations listed above in addition
to the following operations.
• Change user accounts
• Change Operator
access to features
• Update system software and
firmware
• Change data management
settings
To change Operator and/or Supervisor feature
restrictions, complete the following steps.
1. Press SETTINGS > RESTRICTIONS and log in.
Supervisors can log in only if the Change Operator
Access checkbox is selected on the SUPERVISOR
ACCESS tab.
2. Press the SUPERVISOR ACCESS tab to select the
features available to supervisors.
3. Press the OPERATOR ACCESS tab to select
features available to Operators.
4. Press SAV E.
Changing password settings
Users can change the password strength requirements
for user passwords, such as:
• Required length (number of characters)
• Minimum number of lower case characters
• Minimum number of UPPER case characters
• Minimum number numeric characters.
• Minimum number of special characters (#, ~, !, etc.)
NOTE: Users have up to five attempts to enter the
correct password. After the fifth incorrect
attempt, the account is locked and an
Administrator must reset the password in order to
unlock the account.
Administrators cannot reset to a password
previously assigned to the user.
1. Press SETTINGS > RESTRICTIONS > PASSWORDS
and log in.
2. Set passwords to expire at 30 days, 60 days,
90 days, 180 days, or never.
3. Select the Pre-expire passwords on creation
checkbox to make users change their password the
first time they log in.
4. Select the Allow deletion of users checkbox to
delete user information when a user is removed
from the active list.
5. Select the password strength options.
6. Press SAVE.