OsmoTECH PRO Multi-Sample Micro-Osmometer Service Guide
Appendix H: Known Limitations
Additives such as anticoagulants from collection tubes may contribute to the measured osmolality.
Particulate matter can cause premature crystallization, such as sample pre-freeze. The particulate matter may be
eliminated by centrifugation or by filtration.
Samples with air bubbles do not freeze reliably and may yield inaccurate results.
Samples allowed to evaporate will yield higher osmolality values.
Non-aqueous solutions will not yield accurate results. The OsmoTECH PRO is intended to test aqueous solutions only.
Highly viscous solutions may not freeze reliably and may yield inaccurate results.
High concentrations of ethanol, acetone, methanol, isopropanol, ethylene glycol, diethyl ether, paraldehyde,
trichloroethane, or propylene glycol may not freeze reliably and may yield inaccurate results.