Positioning the wall stand bucky
Inserting a DR detector in the wall stand bucky
Securing the detector in the bucky
Positioning the wall stand bucky
To adjust the height (mechanical brake):
Figure 26: Mechanical brake
1. Release the mechanical brake for vertical movement, by turning the knob
of the vertical movement brake counterclockwise.
The bucky can be moved up and down.
2. Lock the bucky into position, by turning the knob clockwise.
WARNING:
When the brakes are released, the carriage tends to lift upward
with the bucky. This balancing helps the user to vertically
position the bucky. With empty bucky and no additional
accessories mounted, it is necessary to accompany the bucky
along its vertical movement and to avoid violent collision when
reaching the bumper at the upper end of the travel range.
WARNING:
Do not move the bucky with excessive force to the end stop
positions.
To tilt the bucky:
Figure 27: Tilting brake