EasyManuals Logo

Agilent Technologies 1100 Series User Guide

Agilent Technologies 1100 Series
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #132 background imageLoading...
Page #132 background image
132
Running an Injector Program
Saving the Method
Table 4 Injector Program Lines
Line Function Parameters Comment
1 DRAW 2 µl (AMOUNT) from
Air (SOURCE)
Draw 2 µl from air into the needle
capillary, to separate the mobile
phase sitting in the capillary from
the sample.
2 DRAW 1 µl (AMOUNT) from
vial 1(SOURCE)
Draw 1µl from derivatization
reagent from vial #1 in the
autosampler tray into the needle
capillary.
3 WASH in vial 2 (VIAL), 1 time
(Cycles)
Wash the needle tip in the wash
vial #2.
4 DRAW 2 µl (AMOUNT) from
sample (SOURCE)
Add 2 µl of the sample in the
autosampler tray to the
derivatization volume already in
the needle capillary.
5 WASH in vial 2 (VIAL), 1 time
(CYCLE)
Wash the needle tip in the wash
vial #2.
6 DRAW 1 µl (AMOUNT) from
vial 1 (SOURCE)
Add 1 µl of derivatization reagent
from vial #1 in the autosampler
tray to the derivatization reagent
and sample already in the needle
capillary.
7 MIX 6 µl (AMOUNT) in
seat (SOURCE), at
500 µl/min (SPEED), 8
times (REPEAT)
Mix 6 µl with a repetition of 8
times at a speed of 500 µl/minute
in the seat.
8 WAIT 0.5 minutes (WAIT) Wait for 0.5 minute for the
mixture to completely react.
9 INJECT Inject the mixture and start the
analysis.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals