EasyManuals Logo
Home>Agilent Technologies>Security Sensors>1260

Agilent Technologies 1260 User Manual

Agilent Technologies 1260
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
Agilent 1260 Infinity DAD and MWD User Manual 95
Using the Detector
5
Setting up an Analysis
10 Enter the pump parameters mentioned under Table 22 on page 87.
11 Pump the water/acetonitrile (30/70 %) mobile phase through the column for
10 minutes for equilibration.
up to 8 signals (A to H) with individual wavelength
settings can be selected.
spectrum settings, “Spectrum Settings (DAD only)” on
page 108
stop and post time can be set (if required)
depending on the application, the lamps can be
selected (one or both).
peak width depends on the peaks in the chromatogram,
“Peakwidth Settings” on page 110
autobalance to zero absorbance (on the analog output
plus offset) at begin and/or end of run.
mechanical slit width can be changed for further
optimization, “Slit Settings” on page 112
margin for negative absorbance, “Margin for Negative
Absorbance Settings” on page 113
Under More additional diagnostic signals can be added
for troubleshooting purpose, see section “Diagnostic
Signals“ in the Service Manual.
time table for programmable actions during the run.
NOTE: The Agilent G1315C/D and G1365C/D time
table can contain a maximum of 60 rows.

Table of Contents

Other manuals for Agilent Technologies 1260

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 and is the answer not in the manual?

Agilent Technologies 1260 Specifications

General IconGeneral
BrandAgilent Technologies
Model1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals