EasyManua.ls Logo

Agilent Technologies 1260 - Page 95

Agilent Technologies 1260
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1260 Infinity DAD and MWD User Manual 95
Using the Detector
5
Setting up an Analysis
10 Enter the pump parameters mentioned under Table 22 on page 87.
11 Pump the water/acetonitrile (30/70 %) mobile phase through the column for
10 minutes for equilibration.
up to 8 signals (A to H) with individual wavelength
settings can be selected.
spectrum settings, “Spectrum Settings (DAD only)” on
page 108
stop and post time can be set (if required)
depending on the application, the lamps can be
selected (one or both).
peak width depends on the peaks in the chromatogram,
“Peakwidth Settings” on page 110
autobalance to zero absorbance (on the analog output
plus offset) at begin and/or end of run.
mechanical slit width can be changed for further
optimization, “Slit Settings” on page 112
margin for negative absorbance, “Margin for Negative
Absorbance Settings” on page 113
Under More additional diagnostic signals can be added
for troubleshooting purpose, see section “Diagnostic
Signals“ in the Service Manual.
time table for programmable actions during the run.
NOTE: The Agilent G1315C/D and G1365C/D time
table can contain a maximum of 60 rows.

Table of Contents

Other manuals for Agilent Technologies 1260

Related product manuals