ARM:RPM:MODE command/query
Enables the Start RPM Arming qualifier.
Command Syntax: ARM:RPM:MODE OFF|0|UP|DOWN
Example Statements: OUTPUT 711;"Arm:Rpm:Mode UP"
OUTPUT 711;"ARM:RPM:MODE DOWN"
Query Syntax: ARM:RPM:MODE?
Return Format: OFF|UP|DOWN
Attribute Summary: Option: not applicable
Synchronization Required: no
Preset State: UP
SCPI Compliance: instrument-specific
Description:
For a runup measurement send ARM:RPM:MODE UP. The first arm occurs when the RPM value
reaches the specified threshold value (specified with ARM:RPM:THR).
For a rundown measurement send ARM:RPM:MODE DOWN. The first arm occurs when the RPM
value reaches the specified threshold value (specified with ARM:RPM:THR).
To specify subsequent arms to occur at RPM intervals measured from the starting RPM value see the
ARM:TIMer command.
In order analysis instrument mode (INST:SEL ORD; Option 1D1), you can disable the Start RPM
Arming qualifier. Send ARM:RPM:MODE OFF.