To select composite power for the specified channel, send CALC[1|2|3|4]:FEED ‘XFR:POW:
COMP [1|2|3|4]’. Order track must be on (sent with the ORD:TRAC:STAT ON command). The
[SENSe:]WINDow:ORDer:DC command specifies whether the analyzer should use the dc bins in
calculating composite power.
In octave analysis instrument mode (INST:SEL OCT; Option 1D1), available commands are:
CALC[1|2|3|4]:FEED ‘D1|D2|D3|D4|D5|D6|D7|D8’.
CALC[1|2|3|4]:FEED ‘TCAP [1|2|3|4]’.
CALC[1|2|3|4]:FEED ‘W1|W2|W3|W4|W5|W6|W7|W8’.
CALC[1|2|3|4]:FEED ‘XFR:POW [1|2|3|4]’.
In swept sine instrument mode (INST:SEL SINE; Option 1D2), available commands are:
CALC[1|2|3|4]:FEED ‘D1|D2|D3|D4|D5|D6|D7|D8’.
CALC[1|2|3|4]:FEED ‘W1|W2|W3|W4|W5|W6|W7|W8’.
CALC[1|2|3|4]:FEED ‘XFR:POW:LIN [1|2|3|4]’.
CALC[1|2|3|4]:FEED ‘XTIM:VOLT [1|2|3|4]’.
CALC[1|2|3|4]:FEED ‘XFR:POW:CROS [1,2|1,3|1,4|3,4]’. The analyzer must be in 2 channel
instrument mode (INPut 2 ON) for selection 1,2. INPut 4 must be ON for selections 1,3 and 1,4.
INPut 4 must be ON and SENse:REFerence must be PAIR for selection 3,4.
CALC[1|2|3|4]:FEED ‘XFR:POW:RAT [2,1|3,1|4,1|4,3]’. The analyzer must be in 2 channel
instrument mode (INPut 2 ON) for selection 2,1. INPut 4 must be ON for selections 3,1 and 4,1.
INPut 4 must be ON and SENse:REFerence must be PAIR for selection 4,3.
To display the normalized variance for the specified channel, send CALC[1|2|3|4]:FEED
‘XFR:POW:VAR [1|2|3|4]’.