CALCulate[1|2|3|4]:LIMit:LOWer:TRACe[:IMMediate] command
Converts the specified trace into a lower limit line.
Command Syntax: CALCulate[1|2|3|4]:LIMit:LOWer:TRACe[:IMMediate]
Example Statements: OUTPUT 711;"calc2:limit:low:trace:imm"
OUTPUT 711;"Calc2:Limit:Low:Trace"
Attribute Summary: Option: not applicable
Synchronization Required: no
Preset State: not applicable
SCPI Compliance: instrument-specific
Description:
CALCulate[1|2|3|4] specifies the trace box. Use CALC1 to specify trace box A, CALC2 to specify B,
CALC3 to specify C, and CALC4 to specify D.
The limit line becomes active for the specified trace box.