CALCulate[1|2|3|4]:CFIT:ABORt command
Aborts the curve fit operation.
Command Syntax: CALCulate[1|2|3|4]:CFIT:ABORt
Example Statements: OUTPUT 711;":Calc3:Cfit:Abort"
OUTPUT 711;"CALC:CFIT:ABOR"
Attribute Summary: Option: 1D3 Curve Fit/Synthesis
Synchronization Required: no
Preset State: not applicable
SCPI Compliance: instrument-specific
Description:
This command aborts the current curve fit operation. As a result of the CALC:CFIT:ABOR command,
the curve fit table is not updated.
Note This command is not trace specific. It ignores the trace specifier.