CALCulate[1|2|3|4]:LIMit:BEEP[:STATe] command/query
Turns the limit-fail beeper on and off.
Command Syntax: CALCulate[1|2|3|4]:LIMit:BEEP[:STATe] OFF|0|ON|1
Example Statements: OUTPUT 711;":Calc2:Lim:Beep ON"
OUTPUT 711;"CALC:LIMIT:BEEP:STAT OFF"
Query Syntax: CALCulate[1|2|3|4]:LIMit:BEEP[:STATe]?
Return Format: Integer
Attribute Summary: Option: not applicable
Synchronization Required: no
Preset State: OFF (+0)
SCPI Compliance: instrument-specific
Description:
The limit-fail beeper emits an audible tone when all of the following conditions are met:
CALC:LIM:BEEP is ON.
CALC:LIM:STAT is ON.
The trace falls outside its current limits.
You can use CALC:LIM:LOW:SEGM and CALC:LIM:UPP:SEGM or CALC:LIM:LOW:TRAC and
CALC:LIM:UPP:TRAC to define a trace’s current limits via the GPIB.
If a trace specifier is not used, the command defaults to trace A.
CALCulate
6-33