CALCulate[1|2|3|4]:CFIT:FREQuency:STARt command/query
Specifies the start frequency for a curve fit operation over a limited frequency span.
Command Syntax: CALCulate[1|2|3|4]:CFIT:FREQuency:STARt {<num
-
ber>[<unit>]}|<step>|<bound>
<number> ::= a real number (NRf data)
limits: 0:114999.9023
<unit> ::= [HZ]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
OUTPUT 711;"CALC4:CFIT:FREQ:START 10.2"
OUTPUT 711;"calc3:cfit:frequency:star 600"
Query Syntax: CALCulate[1|2|3|4]:CFIT:FREQuency:STARt?
Return Format: Real
Attribute Summary: Option: 1D3 Curve Fit/Synthesis
Synchronization Required: no
Preset State: +0.00
SCPI Compliance: instrument-specific
Description:
This command is not valid unless the CALC:CFIT:FREQ OFF command has been sent.
Note This command is not trace specific. It ignores the trace specifier.