CALCulate[1|2|3|4]:CFIT:WEIGht:REGister command/query
Selects the data register which contains the weighting function for the curve fit operation.
Command Syntax: CALCulate[1|2|3|4]:CFIT:WEIGht:REGister
D1|D2|D3|D4|D5|D6|D7|D8
Example Statements: OUTPUT 711;":CALC:CFIT:WEIG:REGISTER D7"
OUTPUT 711;"calc:cfit:weight:reg D1"
Query Syntax: CALCulate[1|2|3|4]:CFIT:WEIGht:REGister?
Return Format: D1|D2|D3|D4|D5|D6|D7|D8
Attribute Summary: Option: 1D3 Curve Fit/Synthesis
Synchronization Required: no
Preset State: D7
SCPI Compliance: instrument-specific
Description:
The default register is D7.
If you select the auto weight feature (the default is CALC:CFIT:WEIG:AUTO ON), the analyzer
automatically generates a weighting function and stores it in the register specified with this command.
If you disable the auto weight feature by sending the CALC:CFIT:WEIG:AUTO OFF command, the
curve fit operation uses the weighting function stored in the register specified with this command. The
analyzer does not generate and store a new weighting function.
The curve fit operation will abort if the curve fit weight register is empty or contains invalid data.
Note This command is not trace specific. It ignores the trace specifier.
CALCulate
6-19

Table of Contents

Other manuals for Agilent Technologies 35670A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 35670A and is the answer not in the manual?

Agilent Technologies 35670A Specifications

General IconGeneral
BrandAgilent Technologies
Model35670A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals